Medical Engineering & Physics
Volume 31, Issue 3 , Pages 314-319, April 2009

Analysis of stresses in complete upper dentures with flat teeth at differing inclinations

  • Anthony E. Prombonas

      Affiliations

    • Department of Dental Technology, Technological and Educational Institute of Athens, Greece
    • Corresponding Author InformationCorresponding author at: Skiathou 44, Glyfada, Athens 16561, Greece. Tel.: +30 1 9634492.
  • ,
  • Dimitris S. Vlissidis

      Affiliations

    • Department of Dentistry, University of Athens, Greece

Received 30 August 2007; received in revised form 8 June 2008; accepted 22 June 2008. published online 04 August 2008.

Abstract 

This study measures the effect of the inclination of the occlusal surfaces of flat teeth, on the magnitude of exerted stresses in a model of the complete upper denture (CUD), in order to estimate which inclination could be more beneficial for the stressed denture.

Three groups of dentures were produced, with three different inclinations of their posterior flat teeth on the buccolingual direction: group 1 with an inward inclination of flat posterior teeth on the buccolingual direction (anti-Monson arrangement), group 2 with flat occlusion (no inclination) and group 3 with outward inclination (Monson arrangement). Using commercial edentulous molds and standardized procedures, six identical CUDs were produced for each group. Two rosette strain gauges were cemented onto the midline of each denture specimen, for measuring the stress magnitudes. The use of flat teeth with outward inclination may prolong the lifetime of the denture, by reducing the detrimental stresses (tensile principal and maximum shear stress). When flat posterior teeth with inward inclination are used, the reinforcement of CUD is necessary, to withstand the significant increase of the developed stresses (p<0.05).

Keywords: Complete upper denture, Flat teeth, Stress pattern, Stress analysis, Denture deformation, Principal stresses, Maximum shear stress, Rosette strain gauge

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

PII: S1350-4533(08)00109-4

doi:10.1016/j.medengphy.2008.06.008

Medical Engineering & Physics
Volume 31, Issue 3 , Pages 314-319, April 2009