Finite element modeling of multi-level cervical spinal segments (C3–C6) and biomechanical analysis of an elastomer-type prosthetic disc

Department of Mechanical Engineering, Hanyang University, Sa-1 Dong, Ansan-City, Kyunggi-Do 425-791, Republic of Korea

Received 2 January 2005; received in revised form 13 September 2005; accepted 28 September 2005. published online 09 November 2005.

Abstract 

A three-dimensional finite element (FE) model for the multi-level lower cervical spinal segment C3–C6 has been developed using computed tomography (CT) data, and applied to study of the effects of the fusion and the artificial disc prosthesis on the biomechanical behavior of the lower cervical spine. The NURBS computer adided dedsig (CAD) data used in this study for modeling the vertebrae facilitate adding surface patch layouts for seamless attachment of the soft tissues, such as intervertebral discs onto the vertebrae. A FE model was completed by generating mesh out of this geometry. Its accuracy was validated by comparing with previously published experimental and numerical results for the flexion-extension, axial rotation, and lateral bending moments. An implantation of an elastomer-type disc prosthesis or fused graft between C4–C5 vertebrae was considered in the FE model by modifying the intact disc. It is shown that the fusion reduced the mobility at its level by about 50–70% for the considered loading cases. It is numerically demonstrated that an elastomer with Young's modulus of 5.9MPa for the artificial disc prosthesis well restores the biomechanical behavior of the intact spine.

Keywords: Lower cervical spine, Finite elements, Artificial disc prosthesis, Fusion

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

PII: S1350-4533(05)00208-0

doi:10.1016/j.medengphy.2005.09.006

Medical Engineering & Physics
Volume 28, Issue 6 , Pages 534-541, July 2006