Medical Engineering & Physics
Volume 28, Issue 2 , Pages 156-165, March 2006

QRS artifact elimination on full night sleep EEG

  • J.-P. Lanquart

      Affiliations

    • Sleep Laboratory, Department of Psychiatry, Erasme Academic Hospital Free University of Brussels, Belgium
    • Corresponding Author InformationCorresponding author. Tel.: +32 2 555 3739; fax: +32 2 555 6955.
  • ,
  • M. Dumont

      Affiliations

    • Biological Physics Laboratory, Université de Mons-Hainaut Mons, Belgium
  • ,
  • P. Linkowski

      Affiliations

    • Sleep Laboratory, Department of Psychiatry, Erasme Academic Hospital Free University of Brussels, Belgium

Received 6 September 2004; received in revised form 18 March 2005; accepted 12 April 2005. published online 06 June 2005.

Abstract 

Spectral analysis is now a standard procedure for analyzing the electroencephalograms (EEG) obtained by polysomnographic recordings. These numerical methods assume an artifact-free EEG since artifacts create spurious spectral components.

Our aim was the development of a QRS artifact removal technique that might be applied to full night EEG with a minimal human intervention. This technique should handle one EEG channel, with or without use of one ECG channel.

Variance minimization, independent component analysis (ICA), morphological filters (MF) have been implemented. Careful attention has been given to define the MF structuring element. The tests on artifact-simulated and real data were checked on the residual ECG spectral components present in the cleaned EEG.

The best results are obtained by the MF when the structuring element is an artifact template defined either directly on the EEG or on the ICA ECG component. Further developments are required to identify and subtract the T-wave artifacts.

Keywords: Sleep EEG, Artifact subtraction, Morphological filters

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

PII: S1350-4533(05)00097-4

doi:10.1016/j.medengphy.2005.04.017

Medical Engineering & Physics
Volume 28, Issue 2 , Pages 156-165, March 2006